XZR200 User’s Manual
6
97337 Issue 4, July 2018
INSTALLATION
3 INSTALLATION
!
When installing the sensor do not use any lubricants or grease
which may contain silicone.
!
The sensor is located in the tip of the probe and is heated to
700°C (1292°F).
Do NOT touch the probe tip with bare hands as this will cause
damage to skin.
3.1
Mounting and Orientation
!
The XZR200 should be grounded (earthed) via the probe as well
as through the cable.
When using the XZR200 in a combustion process the diagram below shows the correct
mounting and orientation. The downward angle of the probe tip will protect the sensor
from dust settling or moisture sitting in the tip and damaging the sensor.
A 12mm
fi
tting with a retaining nut that can be welded to the process and
fi
x the probe
in position. A suitable O-ring should be used to ensure a good seal.
Housing
15°
Partition
Bar Probe
O-Ring
Nut
Weld Socket
Partition
Atmospheric Air
+60°C Max
Gas Mixture
+250°C Max
+400°C Max
Figure 3
Mounting and Orientation
Содержание XZR200
Страница 1: ...XZR200 Oxygen Analyzer User s Manual 97337 Issue 4 July 2018 ...
Страница 24: ...XZR200 User s Manual Michell Instruments 17 CALIBRATION Appendix A Technical Specifications ...
Страница 27: ...XZR200 User s Manual 20 97337 Issue 4 July 2018 APPENDIX B Appendix B Theory of Operation ...
Страница 31: ...XZR200 User s Manual 24 97337 Issue 4 July 2018 APPENDIX C Appendix C Quality Recycling Warranty Information ...
Страница 35: ......
Страница 36: ...http www michell com ...