XZR200 User’s Manual
Michell Instruments
1
INTRODUCTION
1 INTRODUCTION
The XZR200 measures the partial pressure of O
2
utilizing a Zirconium Dioxide (ZrO
2
)
sensor cell in Safe Area (general purpose) applications.
This user manual will show you how to measure oxygen using the XZR200 oxygen
analyzer.
In the following sections you will learn about:
•
Zirconium Dioxide technology
• Analyzer
components
• Installation
• Operation
• Calibration
Please read the manual carefully and pay particular attention to any safety warnings
and noti
fi
cations.
Note: Warnings and items of importance will be marked with bold text.
1.1 Features
•
Accurate linear con
fi
gurable outputs: 4-20mA & 0-10 V DC or RS232
•
Selectable output measurement ranges: Standard ranges of 0 to 25% O
2
and 0 to 100% O
2
or fully adjustable via RS232 when con
fi
gured in 0 to
100% O
2
mode
•
Externally triggered automatic or manual calibration
•
Calibrated in ambient air (20.7% O
2
) or in any other known O
2
concentration
•
Cycling 3.3 V DC logic output allows direct monitoring of the O
2
sensor
pump cycle for diagnostic purposes
• Selectable
output
fi
ltering allows adaptive, fast and dynamic, or slow and
stable output
•
IP65 die-cast aluminum case with stainless steel probe 210mm (8.27") or
400mm (15.75")
Содержание XZR200
Страница 1: ...XZR200 Oxygen Analyzer User s Manual 97337 Issue 4 July 2018 ...
Страница 24: ...XZR200 User s Manual Michell Instruments 17 CALIBRATION Appendix A Technical Specifications ...
Страница 27: ...XZR200 User s Manual 20 97337 Issue 4 July 2018 APPENDIX B Appendix B Theory of Operation ...
Страница 31: ...XZR200 User s Manual 24 97337 Issue 4 July 2018 APPENDIX C Appendix C Quality Recycling Warranty Information ...
Страница 35: ......
Страница 36: ...http www michell com ...