WM281/291 User’s Manual
Michell Instruments
17
appendix B
122
Software Release
R
-
-
6.00
2
-
Reads Software release number
in format X.XX
123 –
124
RESERVED
R
-
-
0
0
-
-
125
Type of Analog
output for
Temperature
R/W
0
130
1
0
-
get/Set Analog output type for
Temperature
0 = 0 – 20mA
1 = 4 – 20mA
128 = 0 – 1V
129 = 0 – 5V
130 = 0 – 10V
126 - 139
RESERVED
R
-
-
0
0
-
-
140
Type of Analog
output for Humidity
R/W
0
130
1
0
-
get/Set Analog output type for
Humidity
0 = 0 – 20mA
1 = 4 – 20mA
128 = 0 – 1V
129 = 0 – 5V
130 = 0 – 10V
141 –
154
RESERVED
R
-
-
0
0
-
-
155
Low Limit of
Analog output for
Temperature
R/W
-30.0
140.0
-30.0
1
°C/°F
get/Set Minimum value of
Analog output for Temperature
156
High Limit of
Analog output for
Temperature
R/W
-30.0
140.0
140.0
1
°C/°F
get/Set Maximum value of
Analog output for Temperature
157
Low Limit of Analog
output for Humidity
R/W
0.0
100.0
0.0
1
%
get/Set Minimum value of
Analog output for Humidity
158
High Limit of Analog
output for Humidity
R/W
0.0
100.0
100.0
1
%
get/Set Maximum value of
Analog output for Humidity
159
RESERVED
R
-
-
0
0
-
-
160
Type of Analog
output for Calculated
Variable
R/W
0
130
1
0
-
get/Set Analog output type for
Calculated Variable
0 = 0 – 20mA
1 = 4 – 20mA
128 = 0 – 1V
129 = 0 – 5V
130 = 0 – 10V
161 –
174
RESERVED
R
-
-
0
0
-
-
175
Low Limit of Analog
output for Dew Point
R/W
-
-
-
0
-
get/Set Minimum value of Dew
Point output
176
High Limit of Analog
output for Dew Point
R/W
-
-
-
0
-
get/Set Maximum value of Dew
Point output
177 –
179
RESERVED
R
-
-
0
0
-
-
180
Low Limit of Analog
output for Mixing
Ratio
R/W
-
-
-
0
-
get/Set Minimum value of
Mixing Ratio output
181
High Limit of Analog
output for Mixing
Ratio
R/W
-
-
-
0
-
get/Set Maximum value of
Mixing Ratio output
182
Low Limit of Analog
output for Absolute
Humidity
R/W
-
-
-
0
-
get/Set Minimum value of
Absolute Humidity output
183
High Limit of Analog
output for Absolute
Humidity
R/W
-
-
-
0
-
get/Set Maximum value of
Absolute Humidity output
184
Low Limit of Analog
output for Enthalpy
R/W
-
-
-
0
-
get/Set Minimum value of
Enthalpy output
185
High Limit of Analog
output for Enthalpy
R/W
-
-
-
0
-
get/Set Maximum value of
Enthalpy output
186
Low Limit of Analog
output for Frost Point
R/W
-
-
-
0
-
get/Set Minimum value of Frost
Point output
Содержание WM281
Страница 20: ...WM281 291 User s Manual Michell Instruments 13 appendix a Appendix A Technical Specifications...
Страница 22: ...WM281 291 User s Manual Michell Instruments 15 appendix B Appendix B Modbus Holding Register Map...
Страница 28: ...WM281 291 User s Manual Michell Instruments 21 appendix D Appendix D EC Declaration of Conformity...
Страница 29: ...WM281 291 User s Manual 22 97463 Issue 1 July 2014 appendix D Appendix D EC Declaration of Conformity...
Страница 30: ...WM281 291 User s Manual Michell Instruments 23 appendix E Appendix E Quality Recycling Warranty Information...
Страница 42: ...http www michell com...