![Michell Instruments WM281 Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/michell-instruments/wm281/wm281_user-manual_1784697012.webp)
WM281/291 User’s Manual
Michell Instruments
5
INstallatIoN
2.5
Connections
2.5.1 Electrical Connection
Applicable for the following products:
WM281/291
Pin
1
Power
2
Power supply -
3
RS485 system 0v
4
ground
5
Output Channel 1 Tempe
6
Output Channel 1 Temperature -
7
Output Channel 2 RH +
8
Output Channel 2 RH -
9
RS485 Data +
10 RS485 Data -
13 Output Channel 3 Calc
14 Output Channel 3 Calculated -
Do not connect Pin 2 (V-) to
Pin 4 (Ground)
15 - 27 V AC and 18 - 38 V DC
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Figure 3
Electrical Connections
The electrical connection box is located in the main body of the WM281/291 Transmitters.
Undo the four screws on the face of the transmitter and carefully lift up the top panel.
The connector is located on the circuit board.
2.5.2 Earth Connection
A suitable shielded conductor cable with a diameter between 5 and 10mm should be
used for entry and exiting the transmitter housing via the 2 x Pg11 metal glands.
The shielded cable must be suitably grounded to ensure electromagnetic protection.
2.5.3 Power Supply
The power supply required to power the system is:
•
18 V to 38 V DC
•
15 to 27 V AC
The DC and AC power supply enter via the same pcb connection pins.
Содержание WM281
Страница 20: ...WM281 291 User s Manual Michell Instruments 13 appendix a Appendix A Technical Specifications...
Страница 22: ...WM281 291 User s Manual Michell Instruments 15 appendix B Appendix B Modbus Holding Register Map...
Страница 28: ...WM281 291 User s Manual Michell Instruments 21 appendix D Appendix D EC Declaration of Conformity...
Страница 29: ...WM281 291 User s Manual 22 97463 Issue 1 July 2014 appendix D Appendix D EC Declaration of Conformity...
Страница 30: ...WM281 291 User s Manual Michell Instruments 23 appendix E Appendix E Quality Recycling Warranty Information...
Страница 42: ...http www michell com...