MAINTENANCE
5
MAINTENANCE
5.1
Replacing the Interchangeable I7000 Module
Figure 4
I7000 Sensor Pin Connector
The following steps should be taken to carry out the replacement of the I7000 module:
1.
Switch off the power supply to the transmitter.
2.
Unscrew the protection filter mounted on the measurement probe body.
NOTE: Be careful not to apply any force to the filter protection
cage.
3.
Taking care not to bend the male pins on the sensor electrical connector,
remove the I7000 to be replaced and insert the new component.
4.
Ensure that the appropriate adhesive label (with the date of calibration)
is attached to the transmitter housing.
5.
Carefully screw the protection filter back on.
Michell Instruments delivers each interchangeable module
with an identification label which must be attached to the
transmitter housing over the previous label.
If a transmitter is returned to the factory for repair
or maintenance without the appropriate label of the
corresponding interchangeable module, the warranty shall be
considered as invalid.
Содержание WM281
Страница 20: ...WM281 291 User s Manual Michell Instruments 13 appendix a Appendix A Technical Specifications...
Страница 22: ...WM281 291 User s Manual Michell Instruments 15 appendix B Appendix B Modbus Holding Register Map...
Страница 28: ...WM281 291 User s Manual Michell Instruments 21 appendix D Appendix D EC Declaration of Conformity...
Страница 29: ...WM281 291 User s Manual 22 97463 Issue 1 July 2014 appendix D Appendix D EC Declaration of Conformity...
Страница 30: ...WM281 291 User s Manual Michell Instruments 23 appendix E Appendix E Quality Recycling Warranty Information...
Страница 42: ...http www michell com...