WM281/291 User’s Manual
16
97463 Issue 1, July 2014
appendix B
Appendix B
Modbus Holding Register Map
B.1
Current Mapping of Modbus Holding Registers
Address
Parameter
Access
Type
Min
Value
Max
Value
Default
Decimal
Points
Unit
Description
0
Temperature Reading
R
-
-
-
2
°C/°F
Returns present value of
temperature channel
1
Humidity Reading
R
0.0
100.0
-
1
%
Returns present value of
humidity channel
2
RESERVED
R
-
-
0
0
-
-
3
Model Type
R
-
-
-
0
-
Returns device model type value
4
RESERVED
R
-
-
0
0
-
-
5
Calculated Variable
Reading
R
-
-
-
1
-
Returns present value of
calculated channel
6-7
Current input
Reading
R
96
480
-
2
mV
Returns the voltage measured
across the main connector
pin which corresponds to the
current input supplied
8– 9
RESERVED
R
-
-
0
0
-
-
10
Actual Low
Temperature Scale
R
-
-
-30.0
1
°C
Returns fixed value -30°C
11
Actual High
Temperature Scale
R
-
-
140.0
1
°C
Returns fixed value 140°C
14
Board Type
R
-
-
0x0430
or
0x0C08
0
-
Returns board type identifier
Read 0x0430 on MS430 board &
Read 0x0C08 on HC08 board
15 – 22
RESERVED
R
-
-
0
0
-
-
23
Temperature Offset
Correction (Single
Point Calibration)
R/W
-999
999
0
2
°C
get/Set Single Point Calibration
offset for Temperature input
24
Digital Temperature
Filter
R/W
0.0
20.0
0.8
1
Seconds
get/Set Digital Filter value for
Temperature input
40 – 44
RESERVED
R
-
-
0
0
-
-
45
Baud rate
R/W
0
4
4
0
-
get/Set COM Baud rate setting
0 = 1200, 1 = 2400, 2 = 4800,
3 = 9600, 4 = 19200
46
ModBus Slave ID
R
0
247
-
0
-
Returns ModBus Slave ID of
device
47
Parity
R/W
0
2
0
0
-
get/Set COM Parity setting
0 = None, 1 = Odd, 2 = Even
48 – 49
RESERVED
R
-
-
0
0
-
-
50
Temperature Open/
Under range Counter
value
R
0
65535
0
0
-
Returns count value for number
of times the device has gone
below Temperature under limit
51
Temperature
Saturate/Over range
Counter value
R
0
65535
0
0
-
Returns count value for number
of times the device has gone
above Temperature over limit
52
Humidity Open/
Under range Counter
value
R
0
65535
0
0
-
Returns count value for number
of times the device has gone
below Humidity under limit
53
Humidity Saturate/
Over range Counter
value
R
0
65535
0
0
-
Returns count value for number
of times the device has gone
above Humidity over limit
57 – 59
RESERVED
R
-
-
0
0
-
-
64
RESERVED
R
-
-
0
0
-
-
69 – 119
RESERVED
R
-
-
0
0
-
-
120
Trademark
R
-
-
5000
0
-
Reads Michell identifier in
ModBus
121
Device Type identifier
R
-
-
129
0
-
Reads type of device
Содержание WM281
Страница 20: ...WM281 291 User s Manual Michell Instruments 13 appendix a Appendix A Technical Specifications...
Страница 22: ...WM281 291 User s Manual Michell Instruments 15 appendix B Appendix B Modbus Holding Register Map...
Страница 28: ...WM281 291 User s Manual Michell Instruments 21 appendix D Appendix D EC Declaration of Conformity...
Страница 29: ...WM281 291 User s Manual 22 97463 Issue 1 July 2014 appendix D Appendix D EC Declaration of Conformity...
Страница 30: ...WM281 291 User s Manual Michell Instruments 23 appendix E Appendix E Quality Recycling Warranty Information...
Страница 42: ...http www michell com...