QMA601 User’s Manual
86
97449 Issue 3, June 2018
APPENDIX C
Register Configuration E — Alarm Configuration
Note: Alarm 4 is a system fault/warning alarm and is configured in register 9
(Configuration M)
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
L4
L3
L2
L1
A3
A3
A3
A3
A2
A2
A2
A2
A1
A1
A1
A1
Alarm1 Parameter (A1)
Alarm2 Parameter (A2)
0000 = Moisture – PPMv (def)
0001= Moisture - PPMw
0010 = Moisture – MGM3
0011 = Moisture – PA (wvp)
0100 = Moisture - LBMMSCF
0101 = Dew point
0110 = Oven temperature
0111 = Flow rate
1000 = Cell Pressure
1001 = Ext Line Pressure
0000 = Moisture – PPMv
0001= Moisture - PPMw
0010 = Moisture – MGM3
0011 = Moisture – PA (wvp)
0100 = Moisture - LBMMSCF
0101 = Dew point
0110 = Oven temperature (def)
0111 = Flow rate
1000 = Cell Pressure
1001 = Ext Line Pressure
Alarm3 Parameter (A3)
Alarm Latch Control (L1 to L4)
0000 = Moisture – PPMv
0001= Moisture - PPMw
0010 = Moisture – MGM3
0011 = Moisture – PA (wvp)
0100 = Moisture - LBMMSCF
0101 = Dew point
0110 = Oven temperature
0111 = Flow rate (def)
1000 = Cell Pressure
1001 = Ext Line Pressure
L1 = 1 = Alarm1 latch
L2 = 1 = Alarm2 latch
L3 = 1 = Alarm3 latch
L4 = 1 = Alarm4 latch
L1 = 0 = Alarm1 don’t latch
L2 = 0 = Alarm2 don’t latch
L3 = 0 = Alarm3 don’t latch
L4 = 0 = Alarm4 don’t latch
Register Configuration F — Analog Output Config
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
T2
T1
O2
O2
O2
O2
O1
O1
O1
O1
Ouput1 Parameter (O1)
Output2 Parameter (O2)
0000 = Moisture – PPMv (def)
0001 = Moisture - PPMw
0010 = Moisture – MGM3
0011 = Moisture – PA
0100 = Moisture - LBMMSCF
0101 = Dew point
0110 = Oven temperature
0111 = Flow rate
1000 = Cell Pressure
1001 = Ext Line Pressure
0000 = Moisture – PPMv
0001= Moisture - PPMw
0010 = Moisture – MGM3
0011 = Moisture – PA
0100 = Moisture - LBMMSCF
0101 = Dew point (def)
0110 = Oven temperature
0111 = Flow rate
1000 = Cell Pressure
1001 = Ext Line Pressure
Output Type – (T1 to T2)
0 = 4–20 mA
1 = 1–5V
(Where T1 is CH1 and T2 is CH2)
Содержание QMA601
Страница 1: ...QMA601 Process Moisture Analyzer User s Manual 97449 Issue 3 June 2018...
Страница 69: ...QMA601 User s Manual Michell Instruments 59 CALIBRATION Figure 45 Typical QMA601 Calibration Certificate...
Страница 80: ...QMA601 User s Manual 70 97449 Issue 3 June 2018 APPENDIX A Appendix A Technical Specifications...
Страница 83: ...QMA601 User s Manual Michell Instruments 73 APPENDIX B Appendix B Hazardous Area Certification...
Страница 86: ...QMA601 User s Manual 76 97449 Issue 3 June 2018 APPENDIX C Appendix C Modbus Register Map...
Страница 105: ...QMA601 User s Manual Michell Instruments 95 APPENDIX D Appendix D Quality Recycling Compliance Warranty Information...
Страница 107: ...QMA601 User s Manual Michell Instruments 97 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 109: ...QMA601 User s Manual Michell Instruments 99 APPENDIX F Appendix F Calculating Conversion Factors for Gas Mixes...
Страница 111: ......
Страница 112: ...http www michell com...