INSTALLATION
2 INSTALLATION
2.1
Analyzer Storage Instructions
In order for this product to be functional upon installation it should be stored in
accordance with the guidelines below:
•
The product must be housed in a sheltered area, out of direct sunlight
and rain.
•
The product should be stored to minimize the possibility of sitting in
ground water.
•
The temperature within the storage environment should be maintained
between -20 to +60°C (-4 to +140°F).
•
The humidity within the storage environment must be non-condensing.
•
The storage environment must not expose the analyzer to any corrosive
elements.
•
The product should stay assembled with its sample conditioning system
(if supplied).
•
All electrical and process connections should remain disconnected and
capped.
•
All protective coatings should remain in place until installation.
•
For prolonged periods of storage, the lid of the packaging crate should be
removed to allow air to circulate.
•
Any documentation supplied with the product should be removed from
the packaging crate and stored elsewhere to protect its integrity.
For the period from installation of the product to commissioning start-up, the following
precautions should be followed:
•
The product and associated sampling system (if supplied) must remain
isolated from the process gas, and the enclosure should remain closed to
ensure ingress protection is maintained.
•
If supplied, the sampling system enclosure heating/thermostat circuit
should be operated if the climatic temperature might fall below +5°C
(+41°F).
•
At time of start-up the procedures contained in the user manuals for both
analyzer and sampling system must be followed.
If the product was previously in service/operation then the following precautions should
be followed before storage:
•
Upon isolation from the gas sample the entire system should be purged
with a dry nitrogen gas before powering down of the analyzer.
•
All connections and ports (gas and electrical) to the analyzer or sample
system (if provided) should be capped.
•
If the product is not removed from its location, the electrical grounding of
the analyzer should remain in place.
Содержание QMA601
Страница 1: ...QMA601 Process Moisture Analyzer User s Manual 97449 Issue 3 June 2018...
Страница 69: ...QMA601 User s Manual Michell Instruments 59 CALIBRATION Figure 45 Typical QMA601 Calibration Certificate...
Страница 80: ...QMA601 User s Manual 70 97449 Issue 3 June 2018 APPENDIX A Appendix A Technical Specifications...
Страница 83: ...QMA601 User s Manual Michell Instruments 73 APPENDIX B Appendix B Hazardous Area Certification...
Страница 86: ...QMA601 User s Manual 76 97449 Issue 3 June 2018 APPENDIX C Appendix C Modbus Register Map...
Страница 105: ...QMA601 User s Manual Michell Instruments 95 APPENDIX D Appendix D Quality Recycling Compliance Warranty Information...
Страница 107: ...QMA601 User s Manual Michell Instruments 97 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 109: ...QMA601 User s Manual Michell Instruments 99 APPENDIX F Appendix F Calculating Conversion Factors for Gas Mixes...
Страница 111: ......
Страница 112: ...http www michell com...