INSTALLATION
The product enclosure is supplied with an external protective earthing/grounding
terminal at the lower right hand side as shown in the figure below. As the first step
of the electrical installation - connect this earthing/grounding terminal to plant earth/
ground by a minimum 4mm
2
earth/ground bond strap. The earthing/grounding terminal
comprises of an earth bolt, 2 x plain washers and 1 x spring washer, which are all nickel
plated.
Figure 3
Earth Bolt And Nut Washer Assembly
Mains Powered Units
As a minimum, the power connection cable should be 3 core over sleeved,
with minimum 0.5mm insulation and rated at 300 V. Cables should have
Live (L), Neutral (N) and Earth [Ground] (E) conductors.
Ensure suitably rated power supply cables and glands are used to ensure that
electrical safety is maintained. Connect each of the Live (L), Neutral (N) and
Earth [Ground] (E) conductors to the similarly marked terminals (L, N, E) on
the Power In connector shown above.
24 V DC Units
The cable should be rated at a minimum of 10 A at 50 V DC with minimum
0.5mm insulation. Connect the +24 V conductor to the terminal
and the 0 V conductor to the terminal marked -.
Before applying power, perform a continuity test to ensure that the power supply cable
and product are effectively connected to the protective earth. A protective earth terminal
is mounted internally and the Earth wire connected to it should never be disconnected.
Do not remove or exchange any of the cables, electrical components or any other parts
supplied with this product. Doing so will invalidate all warranties.
If installing rigid conduit, a stopping fitting must be installed within 46cm (18") of the
enclosure.
There are no additional or special electrical safety requirements other than those
referred to in this manual.
Содержание QMA601
Страница 1: ...QMA601 Process Moisture Analyzer User s Manual 97449 Issue 3 June 2018...
Страница 69: ...QMA601 User s Manual Michell Instruments 59 CALIBRATION Figure 45 Typical QMA601 Calibration Certificate...
Страница 80: ...QMA601 User s Manual 70 97449 Issue 3 June 2018 APPENDIX A Appendix A Technical Specifications...
Страница 83: ...QMA601 User s Manual Michell Instruments 73 APPENDIX B Appendix B Hazardous Area Certification...
Страница 86: ...QMA601 User s Manual 76 97449 Issue 3 June 2018 APPENDIX C Appendix C Modbus Register Map...
Страница 105: ...QMA601 User s Manual Michell Instruments 95 APPENDIX D Appendix D Quality Recycling Compliance Warranty Information...
Страница 107: ...QMA601 User s Manual Michell Instruments 97 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 109: ...QMA601 User s Manual Michell Instruments 99 APPENDIX F Appendix F Calculating Conversion Factors for Gas Mixes...
Страница 111: ......
Страница 112: ...http www michell com...