INSTALLATION
2.4
Mounting the Analyzer
The analyzer is housed in an aluminum Exd enclosure suitable for wall or panel
mounting. Four mounting points are available with M10 clearance holes on fixing
centers of X = 308mm and Y = 312mm.
The enclosure provides environmental ingress protection IP66/NEMA4 and should be
mounted vertically in a location free of any appreciable vibration. It should be placed in
a shaded position to prevent heating effects through sun radiation. The weight of the
analyzer is 35kg (77lbs).
Conduit entries are intended for connection to threaded rigid metal conduit or other
wiring methods in accordance with Article 501 of the National Electrical Code ANSI/
NFPA 70-2005.
4 off M20
threaded entries
External Earth Point
Marking Label
Ø 41
0m
m
(16
.14
”)
335mm (13.2”)
335mm
(13.2”)
308mm
(12.1”)
312mm (12.3”)
4 off M10
41mm
(1.6”)
50mm
(2.0”)
97mm
(3.8”)
57mm (2.2”)
131mm (5.2”)
280mm (11.0”)
41mm
(1.6”)
66mm
(2.6”)
TOP VIEW
FRONT VIEW
RIGHT VIEW
Sample In
Desiccant
Column
Outlet
Enclosure
Breather
(do not obstruct)
Figure 2
Mounting Dimensions
Содержание QMA601
Страница 1: ...QMA601 Process Moisture Analyzer User s Manual 97449 Issue 3 June 2018...
Страница 69: ...QMA601 User s Manual Michell Instruments 59 CALIBRATION Figure 45 Typical QMA601 Calibration Certificate...
Страница 80: ...QMA601 User s Manual 70 97449 Issue 3 June 2018 APPENDIX A Appendix A Technical Specifications...
Страница 83: ...QMA601 User s Manual Michell Instruments 73 APPENDIX B Appendix B Hazardous Area Certification...
Страница 86: ...QMA601 User s Manual 76 97449 Issue 3 June 2018 APPENDIX C Appendix C Modbus Register Map...
Страница 105: ...QMA601 User s Manual Michell Instruments 95 APPENDIX D Appendix D Quality Recycling Compliance Warranty Information...
Страница 107: ...QMA601 User s Manual Michell Instruments 97 APPENDIX E Appendix E Return Document Decontamination Declaration...
Страница 109: ...QMA601 User s Manual Michell Instruments 99 APPENDIX F Appendix F Calculating Conversion Factors for Gas Mixes...
Страница 111: ......
Страница 112: ...http www michell com...