Promet I.S. User’s Manual
46
97221 Issue 5, March 2017
APPENDIX A
Appendix A
Technical Speci
fi
cations
Sensors
Sensor Technology
Michell Ceramic Moisture Sensor
Sensor Version
Easidew PRO I.S.
Measurement Range
-100 to +20°Cdp (-148 to +68°Fdp)
Calibration Range
-100 to +20°Cdp (-148 to +68°Fdp)
Calibration
Traceable to British (NPL) and American (NIST) National Humidity
Standards
Accuracy
Dew point: ±1°C between –60 and +20°Cdp (-76 and +68°Fdp)
Moisture content: ±10% of reading
Dew point: ±2°C between –60.1 and -100°Cdp (-76.18 and -148°Fdp)
Resolution
0.1°C b20 and -100°Cdp (+68 and-148°Fdp)
Analysis Pressure
Up to 45 MPa (450 barg / 5801 psig)
Operating Temperature
-40 to +60°C (-40 to +140°F)
Sample Flow Rate
1 to 5 Nl/min (2.1 to 10.6 scfh)
Optional Pressure Sensor
0–138 barg (other ranges available) Accuracy: ±0.25% FS
Control Unit
Display
Two line 6-digits LED, displaying moisture content / dew point (user
toggle) and analysis pressure
Analog Output
Two 4-20 mA (max load 500
Ω
)
User con
fi
gured for parameter, unit and range
Digital Output
RS485 Modbus RTU
Display Mode
Moisture content (ppm
V
)
Moisture content in natural gas (ppm
V
, lb/MMscf, mg/m³)
Dew point (°C or °F)
Pressure (psig, barg)
Pressure Compensation
Fixed value (user programmed) or dynamic input from optional
pressure sensor
Display Resolution
0.1°Cdp, 0.1°Fdp, 0.1-0.001 ppm
V
ideal gas (adjustable), 0.01 ppm
V
natural gas, 0.01 mg/m³, 0.01 lb/MMscf, 1 psig, 0.1 barg
Alarms
Four alarm relays
Control action and set-point are user-programmable
Two Form C contacts rated 10 A, 240 V AC or 8 A, 24 V DC Non-
inductive load
Two Form A contacts rated 5 A, 240 V AC or 4 A 24 V DC Non-
inductive load
I.S. Barriers
Galvanic isolation type, integrated to Control Unit
Power Supply
85 to 265 V AC, 47/63 Hz or 10 to 72 V DC
10 W max power consumption
Operating Environment
Indoor, safe area, 0 to +50°C (+32 to +122°F) < 90% RH
Interconnection Cable
General instrument type, twisted pair, screened, single pair (two pairs
with pressure sensor)
Enclosure
19” sub rack unit
Dimensions 132 x 483 x 375mm (5 x 19 x 14.75”) (h x w x d)
(100mm (4”) minimum rear clearance depth for cables and vents)
Содержание Promet I.S.
Страница 53: ...Promet I S User s Manual Michell Instruments 45 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...Promet I S User s Manual 48 97221 Issue 5 March 2017 APPENDIX B Appendix B Serial Communications...
Страница 61: ...Promet I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Modbus RTU Comms...
Страница 63: ...Promet I S User s Manual Michell Instruments 55 APPENDIX C...
Страница 64: ...Promet I S User s Manual 56 97221 Issue 5 March 2017 APPENDIX D Appendix D Hazardous Area Certification...
Страница 67: ...Promet I S User s Manual Michell Instruments 59 APPENDIX F Appendix E System Drawings...
Страница 69: ...Promet I S User s Manual Michell Instruments 61 APPENDIX F E 2 FM Approved System Drawing...
Страница 71: ...Promet I S User s Manual Michell Instruments 63 APPENDIX F Appendix F Quality Recycling Warranty Information...
Страница 75: ...Promet I S User s Manual Michell Instruments 67 NOTES...
Страница 76: ......
Страница 77: ......
Страница 78: ...http www michell com...