INTRODUCTION
1.4.2 Power Supply and Input/Output Signal
The terminal blocks for the power supply, signal input, signal output and alarm output
are located on the back panel of the control unit (as shown in
Figure 9).
• Power
Supply
The Promet I.S. requires either 85 to 265 V AC or 10 to 72 V DC power input,
depending on the model ordered.
• Signal
Input
There are two 4-20 mA signal input channels from the dew-point and optional
pressure sensors to the control unit. Both input channels are isolated by
built-in galvanic type I.S. barriers.
• Signal
Output
There are two 4-20 mA linear signal output channels. The output channel 1 is
used for the dew point, moisture or pressure signal, with con
fi
gurable ranges.
Output channel 2 has a
fi
xed pressure signal and cannot be con
fi
gured.
There is one RS485 Modbus RTU digital communication port. Refer to
Appendix B.
• Alarm
Output
There are four alarm relays. Alarms 1 and 2 are Form C contacts rated 10A
/ 240 V AC or 8A / 24 V DC, non-inductive load. Alarms 3 and 4 are Form
A contacts rated 5A / 240 V AC or 4A / 24 V DC, non-inductive load. The
control actions and set points of these four alarms are user-programmable.
A fault alarm with adjustable set points is also included.
Содержание Promet I.S.
Страница 53: ...Promet I S User s Manual Michell Instruments 45 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...Promet I S User s Manual 48 97221 Issue 5 March 2017 APPENDIX B Appendix B Serial Communications...
Страница 61: ...Promet I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Modbus RTU Comms...
Страница 63: ...Promet I S User s Manual Michell Instruments 55 APPENDIX C...
Страница 64: ...Promet I S User s Manual 56 97221 Issue 5 March 2017 APPENDIX D Appendix D Hazardous Area Certification...
Страница 67: ...Promet I S User s Manual Michell Instruments 59 APPENDIX F Appendix E System Drawings...
Страница 69: ...Promet I S User s Manual Michell Instruments 61 APPENDIX F E 2 FM Approved System Drawing...
Страница 71: ...Promet I S User s Manual Michell Instruments 63 APPENDIX F Appendix F Quality Recycling Warranty Information...
Страница 75: ...Promet I S User s Manual Michell Instruments 67 NOTES...
Страница 76: ......
Страница 77: ......
Страница 78: ...http www michell com...