Promet I.S. User’s Manual
36
97221 Issue 5, March 2017
OPERATION
•
For the model with active pressure compensation (
P1/P2
), the
pressure reading value will be automatically converted based on
the new unit selected.
•
For the model without active pressure compensation (
G
), the
manually set pressure value will NOT be automatically converted
based on the new unit selected. The value must be manually
changed.
For example - if the pressure is set at 60 barg - when the
pressure unit is changed to
psiG
, the value of the pressure setting will be
60 psig. This must be manually changed to 870 psig.
NOTE: The pressure fault alarm set point under
FAULT
– fault alarm settings
sub-menu - will also switch to the selected unit. The value will be automatically
converted based on the new unit selected.
3.5.2 Settings for the PPMV I (Ideal Gas) Reading
0. 1 2 3
P P M V _ I
The resolution of
PPMV I
(ideal gas) readings can be adjusted between one decimal
point to three decimal points, based on the user’s preference. It will not change the
resolutions of readings with other units.
a.
Increase or decrease the resolution by pressing the
and
keys.
b.
When the desired resolution is on the display, press the
Select
key. The
display will change to the next setting parameter which is for setting the
scrolling intervals of the unit’s full name on the Main Reading Page.
The
PPMV I
(ideal gas) readings at the Main Reading Page, if selected, will then
change to the set resolution.
3.5.3 Setting the Scrolling Interval for a Unit’s Full Name
At the Main Reading Page, the full description of the currently selected unit will scroll
across the display on the bottom line at set intervals. This interval can be set to the
user’s preference.
1 0. 0
S C R O L L
a.
Increase or decrease the scrolling interval (in seconds) by pressing the
and
keys.
b.
When the desired value is on the display, press the
Select
key. The display
will then change to the next sub-menu
FAULT
– fault alarm settings.
Содержание Promet I.S.
Страница 53: ...Promet I S User s Manual Michell Instruments 45 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...Promet I S User s Manual 48 97221 Issue 5 March 2017 APPENDIX B Appendix B Serial Communications...
Страница 61: ...Promet I S User s Manual Michell Instruments 53 APPENDIX C Appendix C Modbus RTU Comms...
Страница 63: ...Promet I S User s Manual Michell Instruments 55 APPENDIX C...
Страница 64: ...Promet I S User s Manual 56 97221 Issue 5 March 2017 APPENDIX D Appendix D Hazardous Area Certification...
Страница 67: ...Promet I S User s Manual Michell Instruments 59 APPENDIX F Appendix E System Drawings...
Страница 69: ...Promet I S User s Manual Michell Instruments 61 APPENDIX F E 2 FM Approved System Drawing...
Страница 71: ...Promet I S User s Manual Michell Instruments 63 APPENDIX F Appendix F Quality Recycling Warranty Information...
Страница 75: ...Promet I S User s Manual Michell Instruments 67 NOTES...
Страница 76: ......
Страница 77: ......
Страница 78: ...http www michell com...