Easidew DryCheck User’s Manual
16
97108 Issue 13, July 2017
OPERATION
3.4.3 Alarm Set-Point Limit Configuration
P
SET
SET
x4
SET
SET
P
Main display
Main display
Set alarm setpoint lower limit
Set span alarm setpoint upper limit
The following procedure is used to set limits to
which the alarm levels can be set (usually after re-
configuring the instrument’s range for Fahrenheit
readings).
Figure 12 shows the operational key sequence.
1. Press
the
P
key once, the display will read
tECH
.
2. Press
the
SET
key once and the display will
flash between
ConF
and
PinP
.
3. Press
the
key four times and the display
will flash between
ConF
and
GEnn
.
4. Press
the
SET
key once, the display will read
SU-L
.
5. Press
the
key once to display a flashing
4 digit number representing the current
minimum alarm level setting. (The default
setting for the °C range is -100.0).
6. Use
the
or
keys to set the required new
value (e.g. -148.0).
7. Press
the
SET
key to accept the new value.
The display will read
SU-u
.
8. Press
the
key once to display a flashing
4 digit number representing the current
maximum alarm level setting. (The default
setting for the °C range is 020.0)
9. Use
the
or
keys to set the required new
value (e.g. 068.0).
10. Press
the
SET
key to accept the new value,
followed by the
P
key to return to the Main
Display.
Figure 12
Set-up Alarm Set-Point Limits
Содержание Easidew DryCheck
Страница 1: ...Easidew DryCheck OnLine Dew Point Hygrometer User s Manual 97108 Issue 13 July 2017...
Страница 34: ...Easidew DryCheck User s Manual 26 97108 Issue 13 July 2017 APPENDIX A Appendix A Technical Specifications...
Страница 36: ...Easidew DryCheck User s Manual 28 97108 Issue 13 July 2017 APPENDIX B Appendix B System Drawing...
Страница 42: ......
Страница 43: ...http www michell com...