OPERATION
3.4
Operating Temperature / ppm
V
Range
3.4.1 Temperature Range Default
The default temperature unit for the Easidew DryCheck is in degrees Celsius. This is
indicated by the °Cdp LED indicator. The default settings associated with this temperature
scale are as follows:
•
Span -100 to +20°Cdp
•
Lower and upper span limits -100 and +20 (display flashes outside this
range)
•
Minimum alarm set-point -100°Cdp
•
Maximum alarm set-point +20°Cdp
To range the instrument for °F, all the above parameters need
to be changed to their Fahrenheit equivalent values (-148 and
+68°F). It is not sufficient just to change the °F/°C units.
To change the range to Fahrenheit follow the procedures in Section 3.4.2.