INSTALLATION
Figure 6
Monitor Rear Panel Connections
Re-transmission Output
The re-transmission output is current sourcing. Connect the positive output to terminal
14 and the negative output to terminal 13. Use appropriately colored wires eg. red
(positive), black (negative).
Terminal
Wire Color
Signal
Supply Information
1
Blue
0 V (GND)
3
Green
4-20 mA loop current
Default 4-20 mA
4
Red
Transmitter loop supply
(+ve)
+24 V DC w.r.t. terminal 1
7
User de
fi
ned
ALr2 (normally closed)
9
7
8
8
User de
fi
ned
ALr2 (normally open)
9
User de
fi
ned
ALr2 (common)
13
User de
fi
ned
Current loop out (-ve)
Default 4-20 mA
14
User de
fi
ned
Current loop out (+ve)
Default 4-20 mA
16
User de
fi
ned
ALr1 (common)
16
17
17
User de
fi
ned
ALr2 (normally open)
23 (AC Version)
Blue
Power in (neutral)
100 – 240 V, 50/60 Hz
24 (AC Version)
Brown
Power in (live)
100 – 240 V, 50/60 Hz
23 (DC Version)
Black
Negative (-)
0 V
24 (DC Version)
Red
Positive (+)
24 V
NOTE: There are no terminals in positions
5, 6, 10, 11, 12, 15, 18, 19, 20, 21 and 22
Table 3
Summary of Electrical Connections
Содержание Easidew DryCheck