5. Monitoring, warning, error, repair
page 29
ELC PE22B-PE35B V1.0-05.19-GB
Subject to technical alterations
PROFIBUS input data (ELC
→
Master)
Byte-
No.
0 1 2 3
4
5
6
7
8
9
10 11 12 13 14
15
Name
[ELC
Serial
Number]
[Software
Version]
[ELC
Nominal
Power]
[Error
Register]
[Warning
Register]
[Lamp
Status]
[Actual
Power]
[Actual
Voltage]
[Actual
Current]
[Reserved]
↓
Byte
6
Bit
15
14
13
12
11
10
9
8
Name
n/a
n/a
Lamp
Power
Low
Supply
ZVS
Earth
Leakage
Fault
Measure-
ment
Overcurrent
Byte
7
Bit
7
6
5
4
3
2
1
0
Name
Lamp
Voltage
High
Over-
voltage
Bus
Connection
Error
internal ELC
Temperature
High
n/a
Mains
Voltage
Low
DC-Link
Undervoltage
In order to avoid an unwanted reaction of the earth fault control during faultless
operation, the isolation resistance of the lamp cables and all lamp connectors
must be greater than 10 M
Ω
.