Mettler Electronics Corp. — Rev.I_09/23/19
42
from the newly entered time. Failure to do this will cause the
treatment to go on indefinitely.
8.
Notes:
- FM and Burst may not be used simultaneously.
- FM may not be used if two channels are selected.
- When using the “AM” mode, once you turn up the intensity to
the desired level, “Restarting” will appear in a window indicating
that AM is starting.
4.14
TENS, Asymmetrical Biphasic Procedure
1.
Turn on the mains power switch located on the back of the unit by
pressing dot on the switch to the “On” position. Image shows
switch in “Off” position.
2.
Press either “1” or “2” to select channel one or two.
3.
Insert the electrode cable into the channel connector for the
channel that you will be using.
4.
Place the electrodes on the patient.
5.
Insert the patient safety cable into the jack located on the back of
the unit and hand the end with the button to the patient letting
him/her know that they can push this button at any time during
the treatment to stop it.
6.
Use the M Wheel to scroll to any of the parameters that you would
like to change. The parameters and their ranges are listed here for
your reference:
Phase Duration:
20-1,000 µs
Frequency: 1-250
pps
Frequency Modulation: 0-250 pps
Amplitude Modulation: 40, 60, 80, and 100%
Burst frequency:
0-30 bps
Type:
CC
or
CV
Options:
Channel(s) 1 or 2
Please note: Frequency Modulation is not available in 2 channel
mode in the interest of patient comfort.
Содержание Sys Stim 240
Страница 2: ...Mettler Electronics Corp Rev I_09 23 19 2 ...
Страница 16: ...Mettler Electronics Corp Rev I_09 23 19 16 ...
Страница 26: ...Mettler Electronics Corp Rev I_09 23 19 26 ...
Страница 50: ...Mettler Electronics Corp Rev I_09 23 19 50 ...
Страница 58: ...Mettler Electronics Corp Rev I_09 23 19 58 ...