Mettler Electronics Corp. — Rev.I_09/23/19
40
4.12
Microcurrent Procedure
1.
Turn on the mains power switch located on the back of the unit by
pressing dot on the switch to the “On” position. Image shows
switch in “Off” position.
2.
Press either “1” or “2” to select channel one or two.
3.
Insert the electrode cable into the channel connector for the
channel that you will be using.
4.
Place the electrodes on the patient.
5.
Insert the patient safety cable into the jack located on the back of
the unit and hand the end with the button to the patient letting
him/her know that they can push this button at any time during
the treatment to stop it.
6.
Use the M Wheel to scroll to any of the parameters that you would
like to change. The parameters and their ranges are listed here for
your reference:
Polarity:
Positive, Negative or Both
Phase Duration:
1-1,000 ms
Frequency: 0.5-500
pps
Options:
Channel(s) 1 or 2
7.
If all of the parameters are set correctly, position the cursor over
the output window and press “Enter”. Then use the M Wheel by
dragging your finger in a clockwise direction to adjust the intensity
up. You will notice that if two channels are being used, they will go
up simultaneously. The intensity in the digital display will go up
and the green bar will get larger as the intensity increases. Moving
your finger around the M Wheel in a counterclockwise direction
will decrease the intensity. The time will begin to count down until
the treatment is complete. A buzzer will sound and the end-of-
treatment message will appear.
Please note
:
If you change time
during a treatment, press “Enter” once the time is entered to
restart the timer at the new time. The time will then count down
from the newly entered time. Failure to do this will cause the
treatment to go on indefinitely.
8.
Notes:
- Stated polarity is for the red lead wire.
Содержание Sys Stim 240
Страница 2: ...Mettler Electronics Corp Rev I_09 23 19 2 ...
Страница 16: ...Mettler Electronics Corp Rev I_09 23 19 16 ...
Страница 26: ...Mettler Electronics Corp Rev I_09 23 19 26 ...
Страница 50: ...Mettler Electronics Corp Rev I_09 23 19 50 ...
Страница 58: ...Mettler Electronics Corp Rev I_09 23 19 58 ...