![metso automation kajaaniMCAi Скачать руководство пользователя страница 46](http://html.mh-extra.com/html/metso-automation/kajaanimcai/kajaanimcai_installation-operating-and-service-manual_1780221046.webp)
kajaaniMCAi – Installation, Operating & Service
- 11.6 -
W4610201 V2.5 EN
Fig. 11.3. Installing a Vortex cooler to an MCAi-FT sensor.
11.C.4. Installing a Vortex cooler
A Vortex cooler is available as option, order code
A4610977 (Vortex cooler assembly, sensors FT)
and A4611009 (Vortex cooler assembly, sensors F/
FS).
The cooler is installed to keep the sensor’s Electron-
ics Unit sufficiently cool. Cooling is needed if the
sensor is installed in a place where the temperature
inside the Electronics Unit rises to 70°C (158°F) or
higher. The temperature inside the Electronics Unit can
be checked with Communicator-i, S
ELFDIAGNOSTICS
.
The Vortex cooler mounting kit contains the follow-
ing parts:
1. Cooler, with diam. 6mm hose connector
2. Screws M6 x 20, 2 pcs
3. Spring washers M6, 2 pcs
4. Heat-conductive grease
• Installing:
1. Remove the two M6 plug screws from the bottom
plate, next to the connecting pipe.
2. Spread a thin, uniform layer of heat-conductive
grease on the black cooler body, mounting side
(smooth).
3. Place the cooler on the bottom plate so that the
mounting holes are aligned and the outlet pipe
points either upwards or to the side, depending on
the position of the sensor in the tube.
4. Install the screws and washers.
5. Connect an air tube (inside diameter 6 mm, 15/64")
to the hose connector and secure with a hose clamp.
• Cooling:
Maximum volume flow in the air inlet is 230 Nl/min
(8.6 scfm). If no flow meter is used, start the air inlet
carefully and gradually increase the supply until the
required cooling efficiency is obtained.
NOTE: The outlet pipe blows hot air! Make sure that there
are no cables or other objects touching the tube or close
to its outlet end.
Содержание kajaaniMCAi
Страница 1: ...kajaaniMCAi Installation operating service manual W4610201 V2 5 EN for software version 2 5 ...
Страница 59: ...kajaaniMCAi Installation Operating Service A1 W4610201 V2 5 EN App 1 Technical specifications ...
Страница 61: ...kajaaniMCAi Installation Operating Service A2 W4610201 V2 5 EN App 2 Contents of delivery ...
Страница 66: ...kajaaniMCAi Installation Operating Service A4 W4610201 V2 5 EN App 4 Spare parts ...
Страница 68: ...kajaaniMCAi Installation Operating Service A5 W4610201 V2 5 EN App 5 Installation instruction ...
Страница 73: ...kajaaniMCAi Installation Operating Service A6 W4610201 V2 5 EN App 6 Specifications Communicator i ...