The MCAi is less sensitive to fillers than to fibers. In
practice the sensor measured slightly over a half of a
change in filler content. Table 9.3 illustrates the sen-
sor’s sensitivity to different fillers relative to its sensi-
tivity to fibers.
Table 9.3. Sensitivity of MCAi to different fillers.
Filler
MCAi sensitivity
Kaolin
0.61
CaCO
3
0.47
Talc
0.7
TiO
2
0.1
Example: at 3% process consistency the filler con-
tent (100% kaolin) changes by 10 % points, which
corresponds to 0.3% Cs. The MCAi sees this as a (0.61
x 0.3 %) = 0.18 % change, and thus the error will be
0.12%.
The filler content information can be connected to
the isolated analog input Ain1 (on Field Connection
Board) as either a current or voltage signal. If a current
signal will be used, it must be converted into a voltage
by connecting a resistor across the positive and negative
poles of analog input Ain1. The voltage is then derived
from the formula:
U = R x I
where R = resistance and I = current signal value.
Scale the voltage into a filler content value as follows:
1. Go from M
AIN
M
ENU
=> S
ELFDIAGNOSTICS
=> S
PE
-
CIAL
FUNCTIONS
=> F
ILLER
CONTENTS
:
ANALOG
INPUT
(Fig. 9.13).
2. Press [F1&F2] E
DIT
and then type the voltage signal
value and the corresponding filler content for the
low and high limits. You can also increase the
filtering if necessary.
3. Press [F3&F4] S
AVE
. The MCAi will prompt:
“Change scaling of the filler contents?” Press Y
ES
and the MCAi will correct the measured consistency
according to the entered filler content.
Filler content correction can be deactivated and
activated by pressing [F7&F8] T
URN
OFF
and [F7&F8]
T
AKE
INTO
USE
.
F 7
F 6
F 5
F 4
F 8
F 3
F 1
F 2
F 9
F 1 0
F 1 1
F 1 2
F 1 3
F 1 4
F 1 5
F 1 6
i
Fig. 9.13. “Filler contents” display.
FILLER CONTENTS
Connect filler contents signal to the
analog input Ain1.
Low limit
High limit
0.00 V=
0.0 %
5.00 V= 0.0 %
Filtering
=
1.0 s
Filler contents:
0.000 V = 0.0 % Not in use
Edit
Take into
use
9.E. Filler Correction through Analog Input
Содержание kajaaniMCAi
Страница 1: ...kajaaniMCAi Installation operating service manual W4610201 V2 5 EN for software version 2 5 ...
Страница 59: ...kajaaniMCAi Installation Operating Service A1 W4610201 V2 5 EN App 1 Technical specifications ...
Страница 61: ...kajaaniMCAi Installation Operating Service A2 W4610201 V2 5 EN App 2 Contents of delivery ...
Страница 66: ...kajaaniMCAi Installation Operating Service A4 W4610201 V2 5 EN App 4 Spare parts ...
Страница 68: ...kajaaniMCAi Installation Operating Service A5 W4610201 V2 5 EN App 5 Installation instruction ...
Страница 73: ...kajaaniMCAi Installation Operating Service A6 W4610201 V2 5 EN App 6 Specifications Communicator i ...