6.1 Management in the Terminal mode
93
2. Type in
ATR
command.
Select
MCU
.
In the
Transfer
⇒
Send File
menu of
HyperTerminal
utility, specify
the path to a MCU program file (with
.bin
extension) in the
Filename
window; and select the
Xmodem
in the
Protocol
window. Click
Send
.
Estimated duration of the update procedure is 1 minute. If the data
displayed previously on the screen has been restored, the update is com-
pleted successfully.
Note:
if MCU program version is not compatible with versions of the
system FPGA and expansion FPGA microcodes, then you will get a warning
message and the expected version number. In this case, the expected version
should be installed instead of the current version.
Note:
when FPGA microcodes and MCU are updated, it is recommended
to press the hardware restart button with any thin blunt pin. The button is
next to the 12V power supply connector on the side panel.
6.1.2.3
Updating file system
1. Connect the
ETS-1000
to PC via USB (see section 6.1.1).
Check the connection using
AT
command.
2. Type in
ATR
command.
Select
FS image
.
In the
Transfer
⇒
Send File
menu of
HyperTerminal
utility, specify
the path to a MCU program file (with
.bin
extension) in the
Filename
window; and select the
Xmodem
in the
Protocol
window. Click
Send
.
Estimated duration of the update procedure is 5 second. If the data
displayed previously on the screen has been restored, the update is com-
pleted successfully.
ETS-1000. Operations manual
Содержание ETS-1000
Страница 1: ...Ethernet Gigabit Ethernet Tester Analyzer ETS 1000 Operations manual Version 1 0 9 2009 Metrotek ...
Страница 6: ...6 ETS 1000 Operations manual ...
Страница 8: ...8 ETS 1000 Operations manual ...
Страница 10: ...10 ETS 1000 Operations manual ...
Страница 12: ...12 ETS 1000 Operations manual ...
Страница 18: ...18 ETS 1000 Operations manual ...
Страница 96: ...96 ETS 1000 Operations manual ...
Страница 108: ...108 ETS 1000 Operations manual ...