METROTEC
42
12 Specifications
Measuring range
100 % to 10
-31
bar O
2
Ambient temperature
0 to 45 degrees Celsius
Measuring accuracy
+/- 0.3 mV of the sensor e.m.f.
+/- 2 degrees Celsius
+/- 2% of the mA output
+/- 2% of the log oxygen partial pressure
Dimensions
210 x 320 x 80 mm (HxWxD)
330 x 320 x 80 with handle
Weight
3.5 kg
Electromagnetic Compatibility
The device complies with the European
Directive
89/336EC.
The
following
generic standards will be satisfied:
Emitted
interference
EN
50081-1
Interference immunity
EN 50082-2
The device can be operated without
restrictions in living and industrial
environments.
Supply voltage
90 - 230 VAC
Heating-up time for sensors
10 to 15 minutes
Reaction speed
ca. 2 seconds
Heating-up output
ca. 200 VA
Control output
ca. 40 VA
Temperature of measuring cell
700
o
C
Temperature measurement
Thermal element Pt 10Rh-Pt
Sample gas temperature
max. 50
o
C
Amount of sample gas
min. 30 l/h
Humidity
Dew point must be filtered out
Warning! Condensate formation
Dust
Dust must be filtered out
Filter out solid matter particles larger than
5
m
2 Analog output
0/4 – 20 mA configurable, floating
3 Relay output
Configurable for alarms, 1A, 24 V (ohmic)
Interface
Ethernet RJ45 socket, USB 2.0
Содержание GSM Series
Страница 6: ...METROTEC 6 ...
Страница 10: ...METROTEC 10 4 3 Drawing of Cable Routing ...
Страница 32: ...METROTEC 32 8 1 Enabling Configuration 1 Select level 2 touch ENTER 1 Input CODE 2 Store ...