METROTEC
30
076 Measured value correction
Menu line, Menu title, [actual value]
Access code
Detailed explanation found in
“Corrections of Measured
Value”
Access with CODE = 5678
No.
Menu Line
Comment
099
>21% oxygen: correction
of
1st analog input
(standard input)
If measured values between 21 and 100 % O
2
are
adjusted, this value must be corrected. The
correction does NOT affect the measured values
below 21% O
2
100
> 21% oxygen: correction
of 2nd analog input
If measured values between 21 and 100% O
2
are
adjusted, this value must be corrected. The
correction does NOT affect the measured values
below 21% O
2
101
Damping of all measured
values
Damping of analog input signals
102
Sample gas selection
Adjustment of flow display for gases with varying
heat transfer
1 = air, nitrogen
2 = argon, argon mixtures
103 Password
Menu line, Menu title, [actual value]
Access code
Access with CODE = 5678
Can be changed by operator
No.
Menu Line
Comment
103
Password
Password for enabling Level 1.
Important entries are protected from unauthorized
changes by hiding secured menus. They can only
be called up after entering the password (see
“Enabling Configuration”).
On this page a new password can be