5.4.4 Specific Information
In case a condensate separator is used, it must be ensured that the collecting tank is
placed at the lowest point of the entire line system, particularly in the case of water.
The dead volume of condensate tanks and filters will likely cause a delay in the
measuring process.
5.4.5 Filter System: Structure
The gas treatment system must be adjusted to this particular task.
A suggestion for a standard system is as follows:
1. Installing an upstream water separator, possibly with automatic condensate
draining.
2. Coarse filter for separating particles with a size over 50
m
.
(Use only if large amounts of dust are present.)
3. Fine filter for separating particles with a size over 5
m
.
It would be an asset if this filter immediately closes and interrupts the flow of
sample gas as soon as it steams up.
5.5 Switching the Measuring Device off
It is advisable to keep the device continuously in operation. This will avoid the
condensation of the steam in heated sensors which may cause corrosion.
Should it become necessary to switch off the device, press the power switch and also
separate the plug from the mains, as required. Please make sure that the measuring
device is flushed with air before deactivating it.
Switching the measuring device off for a short time:
If the device is switched on again after a short time, please note the information in
item “Switching the Measuring Device on”.
6
Selecting the Measured Value
Selecting the Measured Value enables you to call up various views of the measured
partial pressure of oxygen in the display.