3.2 Remote control commands
784 KFP Titrino
83
3.2.2.76. Assembly.Bur.Rates.Forward.Selected
digital, analog
Assembly.Bur.Rates.Forward.Digital
0...150, max.
Assembly.Bur.Rates.Reverse.Selected
digital, analog
Assembly.Bur.Rates.Reverse.Digital
0...150, max.
Expel and aspirating rate.
Digital or analog control. With digital control, the inputted value applies (in
mL/min). "max." means maximum possible rate with the Exchange Unit in
current use.
Analog means rate control with the analog potentiometer on Titrino.
3.2.2.77. Assembly.Bur.Fill
$G, $H, $C
$G starts the 'FILL' mode of the burette function.
3.2.2.78. Assembly.Bur.ModeDis
$G, $S, $H, $C
Assembly.Bur.ModeDis.Selected
volume, time
Assembly.Bur.ModeDis.V
0.0001...0.1...9999
Assembly.Bur.ModeDis.Time
0.25...1...86400
Assembly.Bur.ModeDis.VStop
0.0001...9999, OFF
Assembly.Bur.ModeDis.AutoFill
ON, OFF
Dispensing mode with parameters. The dispensing mode can only be started
and stopped via the RS Control. During a running dosification, no method can
be started at the Titrino.
.Selected: Dispensing of volume increments or during a preset time.
.Volume, .Time: Size of the volume increments or entry of time.
.VStop:
Limit volume for the dispensing.
.AutoFill:
ON means automatic filling after every dispensing.
3.2.2.79. Assembly.Meas.Status
ON, OFF
Assembly.Meas.MeasInput
Ipol, Upol
Assembly.Meas.Ipol
±127...1...+127
Assembly.Meas.Upol
±1270...400...+1270
Measurement in assembly. The measuring function can only be started via
RS Control. When the measuring function is switched on, no method can be
started at the Titrino.
.Input:
Selection of the "measuring input" for polarized electrodes.
.Ipol:
Polarization current in
µ
A.
.Upol:
Polarization potential in mV, entry in steps of 10 mV.
Содержание Titrino 784 KFP
Страница 4: ...Overview 784 KFP Titrino 1 784 KFP Titrino Instructions for Use...
Страница 5: ...Overview 784 KFP Titrino 2 1 Overview Front view of instrument 4 1 3 2 5...
Страница 99: ...3 3 Properties of the RS 232 Interface 784 KFP Titrino 96...
Страница 130: ...6 2 Pin assignment of the Remote socket 784 KFP Titrino 127 6 2 1 Lines of the Remote socket during the titration...