![Metrohm Titrino 784 KFP Скачать руководство пользователя страница 118](http://html1.mh-extra.com/html/metrohm/titrino-784-kfp/titrino-784-kfp_instructions-for-use-manual_1779647118.webp)
5.1 Setting up and connecting the instruments
784 KFP Titrino
115
5.1.4 Connection of a KF Oven
It is expedient to place the oven on 6.2041.180 instrument bridge.
Take care that the gas outlet of the oven enters the titration vessel as
directly as is possible to prevent the formation of condensed water in
the outlet tubing.
Connection of both RS interfaces (cable 6.2125.110) is only neces-
sary when you require the oven results in the 784 Titrino report.
Make sure there is no report output from the oven!
When the RS interfaces are not connected then the following setting
must be made at the Titrino: <PARAM>, >preselections, "Oven:
no".
Oven on 6.2041.180
instrument bridge
6.2141.010 Cable (Remote)
6.2125.110 Cable (RS
768
A
B
C
D
C
B
A
D
784
Содержание Titrino 784 KFP
Страница 4: ...Overview 784 KFP Titrino 1 784 KFP Titrino Instructions for Use...
Страница 5: ...Overview 784 KFP Titrino 2 1 Overview Front view of instrument 4 1 3 2 5...
Страница 99: ...3 3 Properties of the RS 232 Interface 784 KFP Titrino 96...
Страница 130: ...6 2 Pin assignment of the Remote socket 784 KFP Titrino 127 6 2 1 Lines of the Remote socket during the titration...