![Metrohm Titrino 784 KFP Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/metrohm/titrino-784-kfp/titrino-784-kfp_instructions-for-use-manual_1779647051.webp)
3.1 General rules
784 KFP Titrino
48
3.1.3 Status messages
In order to have an efficient control by an external control device, it must also
be possible to query status conditions; they provide information on the status
of the Titrino. The trigger $D initiates output of the status. Status messages
consist of the global status, the detailed status and eventual error messages,
e.g. $S.Mode.KFT;E26. The global status informs on the activity of the proc-
ess, while the detailed status conditions show the exact activity within the
process.
The following global status conditions are possible:
$G
Go:
The Titrino is executing the last command.
$H
Hold:
The Titrino has been held ($H, key <meas/hold> or by an error which
effects the hold status).
$C
Continue:
The Titrino has been restarted actively after hold.
$R
Ready:
The Titrino has executed the last command and is ready.
$S
Stop:
A process has been aborted in an "unnatural manner". e.g. stopped or aborted
because there was an error.
Detailed status conditions
Status conditions of the global $G:
$G .Mode.KFT .Inac
:
Instrument at the beginning or at the end of a titration.
.Req .Id1
:
Instrument in the KFT mode, requesting Id1 after start.
.Id2
:
Instrument in the KFT mode, requesting Id2 after start.
.Id3
:
Instrument in the KFT mode, requesting Id3 after start.
.Smpl
:
Instrument in the KFT mode, requesting sample size after start.
.Unit
:
Instrument in the KFT mode, requesting unit of sample size after start.
.Start
:
Instrument in the KFT mode, processing the start conditions.
.KFT1
:
Instrument in the KFT mode, titrating to the first endpoint.
.Cond.Ok
:
Instrument in the KFT, conditioning, endpoint reached (after the first
startup from the standby mode).
.Cond.Prog
:
Instrument in the KFT mode, conditioning, endpoint not reached
(Conditioning progressing).
$G .Assembly.Bur .Fill
:
Buret in filling process
.ModeDis
:
Buret in DIS mode
Status conditions of the global $H:
The status message of the action which has been held appears.
If the process is held because a monitored limit has been violated, its status
message is $H.Mode.KFT.Titr.
Status conditions of the global $C:
The status conditions of the global $C are identical with the ones of the global
status $G. They appear when the process has been restarted actively from
the status "Hold" ($C, key <meas/hold> or automatically after elimination of
an error).
Status conditions of the global $R:
$R .Mode.KFT.QuickMeas
:
Quick manual measurement from the initial status in mode KFT.
$R .Mode.KFT .Inac
:
Instrument in the KFT mode, inactive.
.Cond.Ok
:
Instrument in the KFT mode, conditioning, endpoint reached.
.Cond.Prog
:
Instrument in the KFT mode, conditioning, endpoint not reached.
$R .Assembly.Bur.ModeDis
:
Buret in the DIS mode, inactive.
Содержание Titrino 784 KFP
Страница 4: ...Overview 784 KFP Titrino 1 784 KFP Titrino Instructions for Use...
Страница 5: ...Overview 784 KFP Titrino 2 1 Overview Front view of instrument 4 1 3 2 5...
Страница 99: ...3 3 Properties of the RS 232 Interface 784 KFP Titrino 96...
Страница 130: ...6 2 Pin assignment of the Remote socket 784 KFP Titrino 127 6 2 1 Lines of the Remote socket during the titration...