3.2 Remote control commands
784 KFP Titrino
72
3.2.2.26. Mode.Parameter.Presel.IReq
id1, id1&2, all, OFF
Mode.Parameter.Presel.SReq
value, unit, all, OFF
Automatic inquiry after the start of the determination. From such an inquiry,
the determination continues if the requested entry/entries is/are made, e.g.
&SmplData.OFFSilo.Id1 (see 3.2.2.52) or with &M $G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.27. Mode.Parameter.Presel.LimSmplSize.Status
ON, OFF
Mode.Parameter.Presel.LimSmplSize.LoLim
0.0...999 999
Mode.Parameter.Presel.LimSmplSize.UpLim
0.0...999 999
Limit control for the sample size.
3.2.2.28. Mode.Parameter.Presel.Oven
COM1, COM2, no
If an oven is connected, its result will be incorporated into the result report of
the Titrino. If there is no oven connected via RS232, the setting of this pa-
rameter has to be "no".
3.2.2.29. Mode.Parameter.Presel.ActPuls
first, all, cond., OFF
Output of a pulse on the I/O line "Activate", see page 126.
3.2.2.30. UserMeth.FreeMem
read only
Memory space, available for user methods or silo lines. $Q sends the number
of free bytes, e.g. "4928".
3.2.2.31. UserMeth.Recall
$G
UserMeth.Recall.Name
up to 8 ASCII characters
UserMeth.Store
$G
UserMeth.Store.Name
up to 8 ASCII characters
UserMeth.Delete
$G
UserMeth.Delete.Name
up to 8 ASCII characters
UserMeth.DelAll
$G
Management of the internal method memory: Load, store and delete methods.
An action is performed if "$G" is sent to the corresponding node just after
entering the name.
Do not use blank characters before and after method name!
.DelAll: Deletes all methods in the user memory.
Содержание Titrino 784 KFP
Страница 4: ...Overview 784 KFP Titrino 1 784 KFP Titrino Instructions for Use...
Страница 5: ...Overview 784 KFP Titrino 2 1 Overview Front view of instrument 4 1 3 2 5...
Страница 99: ...3 3 Properties of the RS 232 Interface 784 KFP Titrino 96...
Страница 130: ...6 2 Pin assignment of the Remote socket 784 KFP Titrino 127 6 2 1 Lines of the Remote socket during the titration...