![Metrohm 2.947.0010 Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/metrohm/2-947-0010/2-947-0010_manual_1779685046.webp)
4.1 Adjusting the lamp settings
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
947 Professional UV/VIS Detector Vario (2.947.00x0)
■
The VIS lamp is installed.
■
The instrument is switched on.
■
The VIS lamp has been burning for at least 30 minutes.
■
The flow-through cell is clean.
■
The flow-through cell is rinsed with ultrapure water.
■
The flow-through cell is free of air bubbles.
■
The intensity spectrum has been checked and shows a cut-off
.
1
Open the
Configuration
program part in MagIC Net.
Double-click on
947 UV/VIS Detector
in the device table.
The
Properties...
window opens.
2
On the
Detector
tab, click on
[Properties...]
to open the detector
settings.
3
In the detector settings, click on
[Adjust automatically]
to start the
automatic lamp adjustment.
The lamp settings are made based on a built-in algorithm. This algo-
rithm calculates and sets the optimized values for
Integration dura-
tion
and
VIS intensity level
. After the automatic adjustment, a
new intensity spectrum is displayed.