4.1 Adjusting the lamp settings
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
947 Professional UV/VIS Detector Vario (2.947.00x0)
Figure 24
Optimized intensity spectrum with UV lamp and VIS
lamp
6
Click on
[OK]
.
The lamp values are transferred to the instrument. The window is
closed. The adjustment is complete.
4.1.2
Operation with UV lamp
NOTICE
Only follow the following steps if the instrument is in operation exclu-
sively with the UV lamp.
The lamp settings must be adjusted if the lamp was replaced or if
checking the intensity spectrum after first start-up displays a cut-off
.
Remove the VIS lamp completely
opening of the lamp with the socket cap. Otherwise the lamp is automati-
cally switched on as soon as the detector settings are opened.
Adjusting the UV lamp and the lamp settings
Prerequisites