3.3 Base tray and bottle holder
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
947 Professional UV/VIS Detector Vario (2.947.00x0)
■
The bottle holder is cleared.
■
All of the cable connections on the rear have been disconnected.
■
The capillaries are removed from the guide ducts between the instru-
ment and the base tray.
■
There are no loose parts in the instrument.
Accessories
■
3 mm hex key (6.2621.100)
1
2
3
Figure 5
Removing the base tray
1
Tilt the instrument sideways and lay it down flat.
2
Loosen the 4 cylinder screws with the 3 mm hex key. Remove the cyl-
inder screws and their washers.
3
Remove the base tray.
Always mount the base tray under the lowermost instrument of the stack.
Mounting the base tray
Prerequisites
■
The instrument is switched off.
■
The bottle holder is cleared.
■
All of the cable connections on the rear have been disconnected.
■
There are no loose parts in the instrument.
■
The instrument is lying on its side, and the bottom surface is visible.
Accessories
■
3 mm hex key (6.2621.100)