4 Operation
873 Biodiesel Rancimat
■■■■■■■■
61
Temperature
Setpoint temperature of the sample.
Range
50 ... 220 °C
Delta T
Temperature correction: indication of how much the block temperature has
to be increased so that the sample reaches the setpoint temperature. This
value is determined through measurement of the deviation by the current
temperature of the sample from the heating block temperature using a cali-
brated, external temperature sensor
.
Note
If no temperature sensor is available for determining Delta T, you can enter
the approximate values listed in the table below, which have been deter-
mined with 5 g silicone oil and a gas flow of 10 L/h (air).
Range
-9.9 ... 9.9 °C