4 Operation
873 Biodiesel Rancimat
■■■■■■■■
147
2 Carry out GLP test with test resistor
■
Click on the menu item
Tools
▶
GLP test
▶
Conductivity...
and
confirm with
[Next >]
.
The following dialog window appears:
■
Select the option
Use test resistor
.
■
Under
Resistance
, enter the value of the test resistor in Ohms
(10000
Ω
for 6.2109.030) and confirm with
[Next >]
.
■
Connect the test resistor to the electrode connector
in the dialog window and confirm with
[Next >]
.
■
Continue with step 4.
3 Carry out GLP test with standard solution
■
Insert a measuring vessel, each filled with 60 mL standard solution,
together with the measuring vessel cover at all measuring position
.