4 Operation
873 Biodiesel Rancimat
■■■■■■■■
123
Calculation active
on | off
(Default value:
off
)
If this option is
activated
, the parameters for formula 3 can be re-entered.
If this option is
deactivated
, the original result is deleted.
Recalculate standard time
on | off
(Default value:
off
)
Activate/Deactivate the recalculation of the
standard time
from the
induc-
tion time
. If the recalculation is deactivated, the original result is kept.
Calculation active
on | off
(Default value:
off
)
If this option is
activated
, the parameters for the calculation of the standard
time can be re-entered.
If this option is
deactivated
, the original result is deleted.
Formula parameters
Factor
Parameter for the calculation of the standard time, which indicates the factor
by which the time changes per 10 °C.
Target temperature
Target temperature for the calculation of the standard time.