4 Operation
873 Biodiesel Rancimat
■■■■■■■■
149
4 Confirm or reject results
✓
Test passed
: The deviation of the measured conductivities
is within the specified tolerance.
✗
Test failed
: The deviation of the measured conductivities is
greater than the specified tolerance.
■
Confirm the results with
[Finish]
.
or
■
Reject the results with
[Cancel]
.
4.8.4.3
Gas flow
For the
GLP test for gas flow measurement
, you require a gas flow
measuring device (not available from Metrohm), which can be used to
measure the gas flow at every channel.
Carrying out the GLP test for the gas flow measurement
Proceed as follows to carry out the GLP test:
1 Select instrument and start test
■
In the
Control window
, use the menu item
View
▶
Unit x
to
select the instrument for which the GLP test is to be carried out.