M3
Page 83
From the PRESET, the reference of temperature is done.
If the temperature increases, the part becomes bigger. With the compensation, the
measurement shall not change, or at least mostly limit the temperature effect,
depending on the accuracy of the defined coefficient.
If you are using the advanced mode, you need to enter the temperature of the
certificate, and the reference temperature.
For example, if you enter a certificate temperature of 20° and a reference temperature
of 30°, when you PRESET, the displayed value will represent the size of the master at
30°.
The temperature function is advanced functionality of M3. It is recommended to make
trials and comparative measurements to validate the process before using it in
production.
The MB-TP module is compatible with PT100 sensors and hot junction isolated
Thermocouple type K. Both of these sensors can be purchased at Metro.