M3
Page 38
Ref Mark, Step and interpolation
. You have to set the correct value for having a
correct measurement.
The step defines the grating period of the probe’s glass scale. The different
values are defined on the following table :
Type of probe
Step
Specto (ST) 12 or 30
20 µm
Metro (MT)12XX or 25XX 2 µm
Metro (MT) 60 or 101
10 µm
Certo (CT)
2 µm
The interpolation defines the division rate of the glass scale, and therefore the
measurement precision :
Example for a probe type Heidenhain Specto ST12 :
The glass scale of this probe is engraved with a step of 20µm, therefore if the
interpolation is set at 200, you will have a resolution of 20/200 = 1/10 µm.
5.3.4. TTL probe with Heindehain pinout
The following information are focused on TTL probes (or Magnescale or
Mitutoyo)
Step, resolution
An indicator inside the
probe allow to re-
calibrate by its own at the
half of the measuring
range when this option is
activated