Doc#
100576
• REV C
(Sept 2017) Page
28 of 42
Example: The configured range is 10 to 90 mils and the current gap is 30 mils.
Bottom Scale Gap = 10 mils
Top Scale Gap = 90 mils
Current = 16 * (30 -10) / (90-10) ) +4mA = 8mA
NOTE: Tolerance is + 0.15 mA.
4.2.1.2 RADIAL VIBRATION
The transmitter current output is linearly proportional to the full scale vibration range
between 4 mA and 20 mA. The following equation represents the relationship between the
vibration and the loop current:
Vibration = (Current (mA) -4) / 16 * (Full Scale Range)
EXAMPLE:
Measured mA
Full Scale Vibration
Actual Vibration
<3.6
5.0 mils, pk-pk
Probe Fault
4.0
5.0 mils, pk-pk
0.0 mils, pk-pk
12.0
5.0 mils, pk-pk
2.5 mils, pk-pk
20.0
5.0 mils, pk-pk
5.0 mils, pk-pk