Doc#
100576
• REV C
(Sept 2017) Page
15 of 42
Figure 11: Configuring by Parameters
3.5.4.1 PROBE SERIES
All DPS units support Metrix MX2030, Metrix 10,000 series 7200 compatible, Bently Nevada
3300, and Bently Nevada 7200 series probes. The DPS also supports one additional probe
type specified at the time of ordering.
3.5.4.2 PROBE DIAMETER
The DPS supports 5mm, 8mm and 11mm probe tip diameters as part of its standard factory-
loaded curve set.
3.5.4.3 TARGET MATERIAL
DPS units support the standard four probe series listed in section 3.5.4.1 calibrated for 4140
steel. The additional probe type specified at the time of ordering may be calibrated to a
variety of materials. Contact Metrix for information on special material calibrations.
3.5.4.4 MEASUREMENT
The Measurement field is only active for the MX2034 transmitter model. The MX2034 sup-
ports three measurements: Radial Vibration, Axial Position, and Shaft Speed (RPM).
3.5.4.4.1 RADIAL VIBRATION
The MX2034 performs a peak to peak detection and drives the 4-20 mA output proportional
to the configured range.
1) Select the new configuration
parameters from the drop down list
2) Click the
Send
button to send the new
configuration to the DPS