Functional Description
62
4-Channel SPO Digital Oscilloscopes
Functional Description
IV -
ACQUISITION System (cont‟d)
« Acquire »
Press the "
Acquire
" button to open the acquisition menu.
Acquisition
Settings
Option
Settings
Description
Acquisition
Normal
This mode is used for an accurate sampling and display
of most signals.
Peak Detect
This mode is used to display noise and to reduce
aliasing
.
High Res
To increase the vertical resolution and to reduce
random noise.
Average
To reduce the random noise on displayed waveforms.
Average: 4, 16, 32, 64,128, 256, 512, 1024
Sinx/x
Sinx
X
Sinusoidal Interpolation "Sinx"
Linear Interpolation "x"
Sequence
Acq Mode
Off
Normal
Single
Display
Mode
Overlay
Waterfall
Frame Set
Setting range :
1 to 80000 (
The max value depends on the time base
coefficient
“S/div” and the number of active channels
)
Mem Depth
7k - 14k -
70k - 140k -
700k - 1.4M
- 7M - 14M
Select the maximum Memory Depth value