Functional Description
22
4-Channel SPO Digital Oscilloscopes
Functional Description
I - VERTICAL
System (cont’d)
Adjusting
vertical
sensitivity V/div
Vertical scale adjusting has « Coarse » and « Fine » mode, vertical sensitivity
range is 2mV/div to 10V/div.
For example, for CH1:
● Pressing “
CH1
”→ “Volts/Div”→ “Coarse”
sets
“V/div” adjust to “coarse”
default value.
Vertical sensitivity can be adjusted in a 1-2-5 sequence from 2mV/div, 5mV/div,
10mV/div to 10V/div.
● Pressing “
CH1
”→“Volts/Div”→ “
Fine
”
sets “V/div” adjust to “fine” This
setting allows a continuous adjustement of the vertical sensitivity between
coarse steps.
Note : Pressing the vertical sensitivity button also toggles the setting from
« Coarse » to « Fine ».
For example : Use the « Fine » vertical sensitivity adjust to determine the channel
bandwidth when the bandwidth limit is « On »:
1° Set the bandwidth limit to 20MHz on CH2
and “Full” bandwidth on CH1
2° Inject on CH1 and CH2 the same
“300mVpp, 1kHz” sinusoïdal signal
3° Adjust the CH1 vertical sensitivity to 50mV/div (coarse) and CH2 (Fine) to
35mV/div (35mV= 0.7x50mV).
Fine adjust
sensitivity
Bandwidth Limit
« 20M » on CH2
21.6MHz
50mV/div
35mV/div