MI 3200 TeraOhm 10 kV
Maintenance
40
Battery cells are stored in the bottom section of the instrument casing under the battery
cover (see
Fig. 31
). In case the batteries become defective please note the following:
Turn the power off and disconnect any measurement accessories or mains
supply cable connected to the instrument before opening the battery cover to
avoid electric shock.
Remove the battery cover.
All six cells have to be replaced and they have to be of the same type.
To easy remove batteries from battery holder first remove the middle batteries
in each columns and then the others (see the Fig. 31).
Insert the batteries correctly (see Fig. 31), otherwise the test instrument will
not operate and battery may be discharged!
To easy insert batteries to battery holder first insert the upper and the lower
batteries in each columns and at end the middle batteries.
The battery cover has to be placed and fixed back.
The Instrument will only work when rechargeable batteries are inside the
instrument
Nominal power supply voltage is 7.2 V DC. Use six NiMH with D cells (dimensions:
diameter = 33 mm, height = 58 mm). See the
Fig. 31
for correct polarity of the batteries.
Fig. 31. Correct polarity of inserted batteries
Ensure batteries are used and disposed of in accordance with Manufacturers guidelines
and in accordance with Local and National Authority guidelines!