MD 9930 Thermal camera
Settings Menu
17
Center
: Measure the center point temperature. Highlight and press
SELECT
to turn
on or off.
Hot
: Capture maximum temperature in the image. Highlight and press
SELECT
to
turn on or off.
Capture:
The hotsnap function. See chapter
Error! Reference source not found.
for more detailed explanation.
Cold
: Capture minimum temperature in the image. Highlight and press
SELECT
to
turn on or off.
Emiss
Set emissivity and reflected temperature for accurate temperature measurements
and enable alarm temperature. Use
UP
and
DOWN
buttons to highlight the menu,
then press
SELECT
to open.
Emiss – Emissivity:
Set Custom by highlighting it, pressing
SELECT
to open, then
use
UP
and
DOWN
to set. The available range is 0.01–1.00 in steps of 0.01. A value
for a material can be chosen in the drop down menu. Navigate to it using
UP
and
DOWN
then press
SELECT
to use.
Temp.comp. – Temperature compensation:
Compensate for any known offset by a
fixed value. Set by selecting, then using
UP
and
DOWN
buttons.
Figure 3: Measure menu