MI 2087 Instaltest 61557
Measurement instructions
23
Notes!
•
In case of present external voltage higher than 9 V a.c./d.c. between test
terminals, the Continuity measurement will not be carried out after pressing
START
key, but the voltage will be displayed, equipped with
“!”
mark! Beep
warning sound will be affected too.
•
If resistance value higher than 5
Ω
(measured with not-compensated instrument)
is displayed, compensation will not be carried out after pressing
CAL
key, but
already effected compensation will be annulled (
C
9
9
9
9
mark will disappear)!
•
If test result is out of measurement range (open test leads),
>1999
Ω
Ω
Ω
Ω
message
will be displayed!
•
bat
message, displayed during or after finishing the measurement means
batteries are too weak to guarantee correct result. Replace the batteries.