MI 3102(H) BT
Eurotest
XE
Measurements
61
PE terminal test procedure
Connect
test cable to the instrument.
Connect
test leads to the item to be tested ( see
).
Touch
PE test probe (the
TEST
key) for at least one second.
If PE terminal is connected to phase voltage the warning message is displayed,
instrument buzzer is activated, and further measurements are disabled in Zloop and
RCD functions.
Warning:
If dangerous voltage is detected on the tested PE terminal, immediately stop all
measurements, find and remove the fault!
Notes:
PE test terminal is active in the INSTALLATION operating mode (except in the
VOLTAGE, Low ohm, Earth and Insulation functions).
PE test terminal does not operate in case the operator’s body is completely insulated
from floor or walls!
For operation of PE test terminal on commanders refer to chapter