MI 3155
Eurotest
XD
Test and measurements
169
(1) peak voltage
(4) Ulim
(2) voltage at disconnection time
(5) moment of disconnection
(3) calculated voltage value
(6) discharging time
Figure 7.103: Discharging time measuring principle
Connection diagram
Figure 7.104: Discharging time measurement
Measurement procedure
Enter the
Discharging time
function.
Set test parameters / limits.
Connect 3-wire test lead to the instrument and on the DUT, see
Connect DUT to the mains supply and Switch it ON, see
Start the measurement.
Measurement will stop automatically when disconnecting DUT from mains supply.
Save results (optional).