MI 3155
Eurotest
XD
Test and measurements
116
7.9
Testing RCDs
Various test and measurements are required for verification of RCD(s) in RCD protected
installations. Measurements are based on the EN 61557-6 standard.
The following measurements and tests (sub-functions) can be performed:
Contact voltage,
Trip-out time,
Trip-out current and
RCD Auto test.
Figure 7.32: RCD menus
Test parameters / limits
I
N
Rated RCD residual current sensitivity
[10 mA, 15 mA, 30 mA, 100 mA,
300 mA, 500 mA, 1000 mA]
I
N/ I
Ndc
Rated RCD residual current sensitivity for special RCDs types
[30 mA / 6 mA d.c.]
1)
RCD type
RCD type
[AC, A, F, B, B+, EV RCD
1)
, MI RCD
1)
]
Use
RCD / PRCD selection
[fixed, PRCD, PRCD-S, PRCD-K, other]
Selectivity
Characteristic
[G, S]
x I
N
Multiplication factor for test current
[0.5, 1, 2, 5]
Phase
Starting polarity
[(+), (-), (+,-)]
Test
Test
[-, L/PE, L1/PE, L2/PE, L3/PE]
Test
Test current shape
[a.c., d.c.]
2)
Uc (P)
Contact voltage, external probe
[On, Off]
Limit Uc
Conventional touch voltage limit
[12 V, 25 V, 50 V]
RCD Standard
Earthing system
1)
Parameter is available only when parameter Use is set to other (for Electrical Vehicle
(EV) RCDs and Mobile installations (MI) RCDs).