MI 3101 EurotestAT: Instrument operation
Automatic testing
29
4.3.5 Storing auto sequence settings (sequence, number, name)
F2
Opens dialog for storing auto sequence settings in auto sequence main menu.
The dialog enables storing existing auto sequence
settings into different location or overwriting existing.
Figure 4.12: Store dialog
Keys:
/
Select the auto sequence number.
TEST
Confirms the storing.
ESC
Returns to auto sequence main menu without changes.
Auto sequence settings are stored in nonvolatile memory. Stored auto sequence
procedures remain in memory until the user changes them.
It is not possible to store any auto sequence in locked location. Locked auto sequence can
be copied in an unlocked location. Stored sequence is unlocked in this case.
Figure 4.13: Store dialog for locked sequence
Figure 4.14: Failed storing
It is possible to unlock all locked sequences (see
4.4.5