GT-526S Manual Rev F
Page 12
To edit pick list items (e.g. Settings Temp
– C/F), press ▲ or ▼ to navigate to the item.
Press Enter to select the item. Press ▲ or ▼ to change the setting. Press ENTER to
save the setting or ESC to cancel and return to the main value.
To edit numeric values (e.g. Count Alarms - Alarm Limit
), press ▲ or ▼ to navigate to
the item. Press Enter to select the item. Press ▲ or ▼ to increment or decrement a
value. Press ◄ or ► to select the next digit. Press ENTER to save the value or ESC to
cancel and return to the main value.
5.2.
Sample Setup Screen
Figure 11 shows the Sample Setup Screen. The 4 parameters are covered in the
following sections.
LOCATION: 001
SAMPLE TIME: 060
HOLD TIME: 000
SAMPLES: SINGLE
Figure 11
– Sample Setup Screen
5.2.1. Location Number
The location number is used to assign a unique number to a location or area. This
important field is included in sample data records (display, printer and serial output).
5.2.2. Sample Time
The sample time determines the amount of time that counts are accumulated while the
pump is running. The length of the sample is user settable from 1
– 999 seconds.
5.2.3. Hold Time
The Hold time is the time between samples when sampling more than a single sample.
The Hold time is user settable from 0
– 999 seconds. The pump will remain on during
the hold period if the Hold time is 60 seconds or less. If the Hold time is greater than 60
seconds, the pump will stop after each sample, and start a few seconds before the next
sample. Hold times greater than 60 seconds will increase pump life.
5.2.4. Samples
The samples setting controls the number of samples to take as illustrated below.
Selection
Description
REPEAT
Repeat configures the unit for continuous sampling.
SINGLE
Single configures the unit for a single sample.
002-999
Configures the unit to take N samples.
5.3.
Settings Screen
Figure 12 shows the Settings Screen. The 4 parameters are covered in the sections
immediately following.